Abstract
Purpose
To quantitatively compare the artefact reduction between standard and metallic artefact reduction (MAR) fast spin echo (FSE) T2 sequences in a low-field magnetic resonance imaging (MRI) scanner (0.3 T) in patients with titanium volar wrist plating.
Materials and methods
Sixteen patients with fractures of the distal radius, treated with titanium volar wrist plating and screws, were examined using a dedicated 0.3 T MRI scanner. Coronal standard FSE T2, FSE T2 high bandwidth (HiBW) and FSE T2 view angle tilting (VAT) sequences were performed. Metallic artefact volume, consisting of both "black" and "bright" artefacts, was calculated for each sequence. Quantitative differences were compared using repeated measures ANOVA test (P < 0.05).
Results
FSE T2 HiBW and FSE T2 VAT showed a significant reduction in artefact volume compared to the standard sequence. Differences between the artefact volume of the standard FSE T2, HiBW and VAT sequences were statistically significant for both the "black" and "bright" artefacts (P < 0.0001). Differences between the 1.5 HiBW and VAT sequences were statistically significant (black P < 0.0001, bright P < 0.0302).
Conclusions
MAR sequences significantly reduced metallic artefacts in vivo using a 0.3 T MRI scanner.