Τρίτη 19 Φεβρουαρίου 2019

Ultra-low dose chest computed tomography: Effect of iterative reconstruction levels on image quality

Publication date: Available online 18 February 2019

Source: European Journal of Radiology

Author(s): Mercy Afadzi, Elisabeth Kirkeby Lysvik, Hilde Kjernlie Andersen, Anne Catrine T. Martinsen

Abstract
Purpose

To optimize image quality and radiation dose of chest CT with respect to various iterative reconstruction levels, detector collimations and body sizes.

Method

A Kyoto Kagaku Lungman with and without extensions was scanned using fixed ultra-low doses of 0.25, 0.49 and 0.74 mGy CTDIvol, and collimations of 40 and 80 mm. Images were reconstructed with the lung kernel, filtered back projection (FBP) and different ASIR-V levels (10 – 100%). Contrast-to-noise ratios (CNR) were calculated for 12 mm simulated lesions of different densities in the lung. Image noise, signal-to-noise ratios (SNR), variations in Hounsfield units (HU), noise power spectrum (NPS) and noise texture deviations (NTD) were evaluated for all reconstructions. NTD was calculated as percentage of pixels outside 3 standard deviations to evaluate IR-specific artefacts.

Results

Compared to the FBP, image noise reduced (5 – 55%) with ASIR-V levels irrespective of dose or collimation. SNR correlated positively (r ≥ 0.925, p ≤ 0.001) with ASIR-V levels at all doses, collimations, and phantom sizes. ASIR-V enhanced the CNR of the lesion with the lowest contrast from 12.7 – 42.1 (0 – 100% ASIR-V) at 0.74 mGy with 40 mm collimation. As expected, higher SNR and CNR were measured in the smaller phantom than the bigger phantom. Uniform HU were observed between FBP and ASIR-V levels at all doses, collimations, and phantom sizes. NPS curves left-shifted towards lower frequencies at increasing levels of ASIR-V irrespective of collimation. A positive correlation (r ≥ 0.946, p ≥ 0.001) was observed between NTD and ASIR-V levels. NTD of the FBP was not significantly (p ≤ 0.087) different from NTD of ASIR-V ≤ 20%. The data from the NPS and NTD indicates a blotchier and coarser noise texture at higher levels of ASIR-V, especially at 100% ASIR-V.

Conclusion

In comparison with the FBP technique, ASIR-V enhanced quantitative image quality parameters at all ultra-low doses tested. Moreover, the use of ASIR-V showed consistency with body size and collimation. Hence, ASIR-V may be useful for improving image quality of chest CT at ultra-low doses.



from #Head and Neck by Sfakianakis via simeraentaxei on Inoreader http://bit.ly/2Ej2x3N

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου

Σημείωση: Μόνο ένα μέλος αυτού του ιστολογίου μπορεί να αναρτήσει σχόλιο.