Objectives The aim of this study was to evaluate if a high-resolution photon-counting detector computed tomography (PCD-CT) system with a 1024×1024 matrix reconstruction can improve the visualization of fine structures in the lungs compared with conventional high-resolution CT (HRCT). Materials and Methods Twenty-two adult patients referred for clinical chest HRCT (mean CTDI vol, 13.58 mGy) underwent additional dose-matched PCD-CT (mean volume CT dose index, 13.37 mGy) after written informed consent. Computed tomography images were reconstructed at a slice thickness of 1.5 mm and an image increment of 1 mm with our routine HRCT reconstruction kernels (B46 and Bv49) at 512 and 1024 matrix sizes for conventional energy-integrating detector (EID) CT scans. For PCD-CT, routine B46 kernel and an additional sharp kernel (Q65, unavailable for EID) images were reconstructed at 1024 matrix size. Two thoracic radiologists compared images from EID and PCD-CT noting the highest level bronchus clearly identified in each lobe of the right lung, and rating bronchial wall conspicuity of third- and fourth-order bronchi. Lung nodules were also compared with the B46/EID/512 images using a 5-point Likert scale. Statistical analysis was performed using a Wilcoxon signed rank test with a P